Scrolls
ANEST IWATA ISP SCROLL PUMP:
As ISP dry scroll vacuum pumps do not contain lubricating oil or sealing water resulting in no back-migration of fluid, it ensures clean vacuum. it is light-weight and occupies minimum space because of original design called double wrap scroll. While orbiting scroll makes one turn, the process of suction, compression and exhaust proceeds continuously, resulting in a minimum pulsation which provides less vibration and low noise. Due to double wrap scroll design which is dynamically stable, ISP dry scroll vacuum pumps can deal with a wide range of applications ranging from atmospheric pressure to the ultimate pressure. There is little power variance between atmospheric pressure and the ultimate pressure, which allows repeated continuous operation between atmospheric pressure and vacuum. The Air Flush device has been added as a standard feature. Shaft seal configuration has also improved. Due to the Air Flush and the new shaft seal system, reliability and maintenance interval as well as evacuating performance of water vapor and small particles have been now remarkably improved.
TECHNICAL DATA | Value | ISP-250B | ISP-500B |
---|---|---|---|
Displacement (Swept Volume)
50 Hz
60 Hz
|
l/min
|
250
300
|
500
600
|
Ultimte Pressure |
Pa (Torr)
|
<1.6 (<1.2 x 10-2)
|
<1 (<7.5 x 10-3)
|
Leak Tightness
|
Pa (mbar)
|
<1.0 x 10-2 (<1.0 x 10-4)
|
<1.0 x 10-2 (<1.0 x 10-4)
|
Maximum Intake Pressure
|
–
|
Atmospheric Pressure
|
Atmospheric Pressure
|
Ambient Operating
Tempurature Range
|
Celcius
|
5 ~ 40
|
5 ~ 40
|
Output
|
kW
|
0.4
|
0.6
|
Inlet Connection
|
–
|
NW25
|
NW40
|
Outlet Connection
|
–
|
NW16
|
NW25
|
Direction of Inlet
|
–
|
Perpendicularly Fitted
|
Perpendicularly Fitted
|
Cooling System
|
–
|
Air-cooled
|
Air-cooled
|
BOC EDWARDS XDS SCROLL PUMP:
The XDS pump is a compact, reliable vacuum pump which is suitable for use on vapour handling processes. The XDS pump may be used for some pumping applications involving corrosive substances and particulates. The body of the pump includes a fixed scroll and an orbiting scroll. The orbiting scroll is driven by the electric motor through an eccentric cam on the motor drive shaft. The movement of the orbiting scroll, meshed with the fixed scroll, forms successive crescent shaped volumes in the pump. Gas which enters the pump through the inlet is compressed by the movement of the orbiting scroll and swept towards the centre of the fixed scroll. The compressed gas enters the exhaust port near the centre of the stationary scroll and is exhausted from the pump through the outlet. The XDS pump is a dry vacuum pump, as all the bearings, with their hydrocarbon lubricant, are isolated from the vacuum space. The inlet of the pump is fitted with a removable inlet strainer which prevents the entry of debris into the pump. The pump mechanism is driven directly by a single phase electric motor. The motor is fitted with an on / off switch (7) and a thermal overload device. The thermal overload device has an automatic reset; when the motor cools down, the device resets and (unless you have incorporated suitable control equipment which must be manually reset), the motor will restart. The pump is air-cooled by a fan mounted at the opposite end to the motor.
TECHNICAL DATA | Value | XDS 5 | XDS 10 | XDS 35i |
---|---|---|---|---|
Ambient Temperature:
Operating Range
Storage Range
|
Celcius
|
10 to 40
-30 to 70
|
10 to 40
-30 to 70
|
10 to 40
-30 to 70
|
Maximum Humidity
|
–
|
90 %
|
90 %
|
90 %
|
Suckback on pump switch
|
mbar
|
30 mbar ls-1
|
30 mbar ls-1
|
30 mbar ls-1
|
Leak tightness
|
mbar
|
1 x 10-6 mbar ls-1
|
1 x 10-6 mbar ls-1
|
1 x 10-6 mbar ls-1
|
Maximum Displacement
50 Hz Supply
60 Hz Supply
|
m3/hr
|
5.7
6.7
|
11.3
13.4
|
43
|
Maximum Pumping Speed
50 Hz Supply
60 Hz Supply
|
m3/hr
|
4.8
6.0
|
9.3
11
|
35
|
Maximum Permitted Inlet Pressure
and Gas-ballast Inlet Pressure
|
–
|
0.5 bar gauge
|
0.5 bar gauge
|
–
|
Maximum Permitted Outlet Pressure
|
–
|
1 bar gauge
|
1 bar gauge
|
–
|
BOC EDWARDS GVSP30 SCROLL PUMP:
TECHNICAL DATA | Value | GVSP30 |
---|---|---|
Ambient Operating Temperature Range
|
Celcius
|
5 to 40
|
Ambient Storage Temperature Range
|
Celcius
|
-20 to 60
|
Ultimate Pressure
|
mbar, Pa
|
1x 10-2mbar, 1 Pa
|
Maximum Inlet Pressure
|
mabr, Pa
|
1000 mabr, 1 x 105Pa
|
Maximum Outlet Pressure
|
mbar, Pa
|
1150 mbar, 1.15 x 105Pa
|
Leak Tightness
|
mbar, Pa
|
< 1 x 10-4mbar l s-1, 1 x 10 Pa l s-1
|
Displacement (Swept Volume)
50 Hz
60 Hz
|
CFM
|
17.4
20.7
|
Peak Pumping Speed
50 Hz
60 Hz
|
CFM
|
14.7
17.6
|
Inlet Connection
|
–
|
NW40
|
Outlet Connection
|
–
|
NW25
|
VARIAN TRI-SCROLL PUMPS:
TECHNICAL DATA | Value | PTS-300 | PTS-600 |
---|---|---|---|
Displacement
50 Hz
60 Hz
|
CFM
m3/hr
|
8.8
12.6
|
17.7
25.2
|
Ultimate Pressure
|
Torr
|
0.01
|
0.007
|
Motor Power
|
Hp
|
.75
|
1
|
Pump Weight
|
lbs
|
57
|
70
|
Inlet Flange
|
–
|
KF16
|
KF40
|
Exhaust Outlet Flange
|
–
|
1/4″ Female NPT
(KF16 adapter included)
|
3/8″ Female NPT
(KF16 adapter included)
|